Table 1 Analytical assay characteristics

From: Comparison of ELISA- and SIMOA-based quantification of plasma Aβ ratios for early detection of cerebral amyloidosis

 

Platform

ELISA colorimetric

SIMOA

Analyte

1–42

1–40

t-tau

1–42

1–40

Assay

Provider

EUROIMMUN

EUROIMMUN

ADx

ADx

ADx

Catalogue number

EQ 6521-9601

EQ 6511-9601

NA

NA

NA

Biofluid

EDTA plasma

EDTA plasma

EDTA plasma

EDTA plasma

EDTA plasma

Status

Commercial

Commercial

Prototype

Prototype

Prototype

Specificity

1–42

1–40

6 tau isoforms

1–42

1–40

Dilution

Pre-dilution factora

4

4

4

4

20

Final sample dilutionb

5

5

5

5.8

29

Calibrator

Type

Recombinant

Recombinant

Recombinant

Recombinant

Recombinant

No. of calibrator points

7

7

7

7

7

Range, pg/mL

1–40

1–75

1–100

1–64

1–64

Patient samples

Number

199

199

199

199

199

Range, pg/mL

8.1–57.1

22.4–311.8

15.0–102.1

7.9–44.0

60.7–160.0

Within cal. range, %

100%

100%

100%

100%

100%

CV%

Intra-assay conc

1.63

1.67

2.15

4.33

2.24

Inter-assay conc

4.82

2.58

13.55

8.59

5.95

Analytical sensitivity

LoD, pg/mL

2.4

5.7

Not determined

0.46

0.77

LoQ, pg/mL

3.5

12.5

Not determined

1.21

1.57

S/N ratio

7.60

65.4

Not determined

11.0

240

Antibodies [26, 27]

Name capture

ADx102 (21F12)

ADx103 (2G3)

ADx203

ADx102 (21F12)

ADx103 (2G3)

Epitope capture (AA)

34–42

33–40

Tau194–204

34–42

33–40

Name detector

2G3

2G3

ADx204

2G3

2G3

Epitope detector (AA)

1–5

1–5

TauN-terminus

1–5

1–5

Assay protocol

Incubation times, h

3–0.5–0.5

3–0.5–0.5

3–0.5–0.5

2–0.08

2–0.08

Incubation T, °C

18–25

18–25

18–25

18–25

18–25

Curve fit

4PL

4PL

4PL

4PL

4PL

QC panel, pg/mL

QC1: high

30.5

189.2

22.3

13.7

83.2

QC2: intermediate

26.6

149.1

18.9

19.7

75.9

QC3: intermediate

22.0

116.8

21.7

13.4

69.7

QC4: low

20.7

107.3

23.7

10.5

33.3

QC5: low spiked

14.4

0.0

NA

21.6

21.8

QC6: high spiked

117.9

125.6

NA

173.7

181.2

C1: low kit control

23.3

114.6

NA

NA

NA

C2: high kit control

45.1

186.9

NA

NA

NA

  1. AA amino acid, β-amyloid, CV coefficient of variation, LoD limit of detection, LoQ limit of quantification, S/N ratio signal to noise ratio, QC quality control, T temperature, t-tau total tau
  2. aSample pre-dilution was performed using assay diluent in polypropylene low-binding 96-well microplates
  3. bFinal sample dilution during sample incubation step in assay protocol